EmStat3 i 3+

EmStat is a small and low cost potentiostat with research grade capabilities. It has eight current ranges from 1 nA to 100 mA (EmStat3+) full scale, with a minimum resolution of 1 pA. It provides the most relevant electroanalytical measurement techniques and is used with the PSTrace software or with PStouch for Android. > Wyślij zapytanie

Specifications

EmStat version: 3 3+ - dc-potential range ± 3.000 V ± 4.000 V - compliance voltage ± 5 V ± 8 V - dc-potential resolution 0.1 mV 0.125 mV - dc-potential accuracy 0.2 % 0.3 % - current ranges 1 nA to 10 mA (8 ranges) 1 nA to 100 mA (9 ranges) - maximum current ± 20 mA typical and ± 15 mA minimum ± 100 mA typical 

The current is measured using a zero resistance ammeter (ZRA).

 - current resolution 0.1% of current range, 1 pA at lowest current range - current accuracy (deviation) ≤ 1 %of current range at 1 nA ≤ 0.5 % at 10 nA ≤ 0.2 % at 100 nA to 100 uA ≤ 0.5 % at 1 mA, 10 mA and 100 mA all with max. 0.2 % offset error - electrometer amplifier input  > 100 Gohm // 4 pF - rise time approx. 100 μs - sensor connection shielded cable with circular connector for WE, RE, CE and Sense (EmStat3+ only) 

General

- housing anodized aluminium: 6.7 x 5.0 x 2.8 cm - weight 85 g - power supply USB 5 V, min. 130 mA (ES3) or 500 mA (ES3+) - communication mini USB - auxiliary port present no

Downloads

EmStat3 Brochure

Kontakt