EmStat is a small and low cost potentiostat with research grade capabilities. It has eight current ranges from 1 nA to 100 mA (EmStat3+) full scale, with a minimum resolution of 1 pA. It provides the most relevant electroanalytical measurement techniques and is used with the PSTrace software or with PStouch for Android.

> Wyślij zapytanie

Specifications

EmStat version: 3 3+
– dc-potential range ± 3.000 V ± 4.000 V
– compliance voltage ± 5 V ± 8 V
– dc-potential resolution 0.1 mV 0.125 mV
– dc-potential accuracy 0.2 % 0.3 %
– current ranges 1 nA to 10 mA (8 ranges) 1 nA to 100 mA (9 ranges)
– maximum current ± 20 mA typical
and ± 15 mA minimum
± 100 mA typical
 

The current is measured using a zero resistance ammeter (ZRA).

 
– current resolution 0.1% of current range, 1 pA at lowest current range
– current accuracy (deviation) ≤ 1 %of current range at 1 nA
≤ 0.5 % at 10 nA
≤ 0.2 % at 100 nA to 100 uA
≤ 0.5 % at 1 mA, 10 mA and 100 mA
all with max. 0.2 % offset error
– electrometer amplifier input       > 100 Gohm // 4 pF
– rise time     approx. 100 μs
– sensor connection shielded cable with circular connector for WE, RE, CE and Sense (EmStat3+ only)
 

General

– housing anodized aluminium: 6.7 x 5.0 x 2.8 cm
– weight 85 g
– power supply USB
5 V, min. 130 mA (ES3) or 500 mA (ES3+)
– communication mini USB
– auxiliary port present no

Downloads

EmStat3 Brochure