Wielokanałowy potencjostat/galwanostat/ZRA

–  zintegrowany FRA/EIS w każdym kanale – 10µHz – 250kHz (opcjonalnie 1 MHz )
–  do 8 modułów/16 kanałów w jednej obudowie – możliwość łączenia do 64 kanałów

> Wyślij zapytanie

Standard channnel specification

Channel Performance:

Standard sModule (2.5A/10V)
4 Electrodes

WE, CE, RE and S
Current compliance

±2.5 A
Maximum output Voltage

±10 V
Potentiostat Bandwidth

>500 kHz
Stability settings

High Speed, Standard, and High Stability
Programmable response filter

1 MHz , 100 kHz , 10 kHz , 1 kHz , 10 Hz
Signal acquisition

dual channel 16 bit ADC, 100.000 samples/sec

Applied potential range

±10 V, with 0.33 mV resolution
Applied potential accuracy

0.2%, or 2 mV
Current ranges

±10 nA to ±10 A in 10 steps
Measured current resolution

0.015% of current range, minimum 1.5 pA
Measured current accuracy


Applied current resolution

0.033% of applied current range
Applied current accuracy

Potential ranges

±10mV, ±100mV, ±1V, ±10V
Measured potential resolution

0.003% of potential range, minimum 400nV
Measured potential accuracy

0.2%, or 2mV
Impedance Analyser

Frequency range

10µHz to 250kHz   (optional: 10µHz to 1MHz)

0.015mV to 1.0V, or 0.03% to 100% of current range
DC offset

16 bit dc offset subtraction, and 2 dc-decoupling filters

Input impedance

>1000 Gohm //<8pF
Input bias current

<10 pA

>5 MHz
Special functions:

Ohmic drop compensation

2 V/current range, 16 bits resolution
Peripheral connections:

(not for dual channel module)
2 Analog in

±10 V, 16 bits resolution, bandwidth 40 kHz
1 Analog out

±10 V, 16 bits resolution
1 Digital input, and 3 Digital outputs

0 to +5 V
I-out, and E-out

analog monitor for cell current and potential

±0.5 V sinewave 10µHz-250kHz with variable attenuation
Channel-X, and Channel-Y inputs

±4 V: to record impedance from peripheral devices


w x d x h = 3 x 35 x 13 cm

0.7 kg
PC requirements

Windows XP/Vista/7/8, with free USB port

Specifications: Ivium-n-Stat main frame

Slot positions

can mount up to 8 modules/ max. 16 channels
Common connectors

GND and combined emergency off control monitor
Power requirements

20A-frame: 100-240 V, 47-63 Hz, 300 W
40A-frame: 100-240V, 47-63Hz, 600W


USB 2.0 and 3 compliant

w x d x h = 47 x 36 x 14 cm

6.2 kg (no modules)11.8 kg (with 8 modules)

Other models specification


sModule sModule dModule dModule
System performance:

5A/10V 2A/10V 500mA/10V 2.5A/10V
Number of channels in module

1 1 2 2
Current compliance

±5A ±2A ±500mA ±2.5A
Maximum output voltage

±10V ±10V ±10V ±10V
Maximum compliance voltage

±10V ±20V ±10V ±10V
Peripheral port

yes yes not available not available


The CompactStat is controlled via Ivium’s own IviumSoft. This versatile, yet intuitive software, allows instrument control, data management and analysis, peripheral instrument control, etc. IviumSoft can also interface with and be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of IviumSoft is standard included with each instrument.

Ivium-n-Stat software control:

  • Control of separate channels individually
  • All channels simultaneously
  • Synchronous start
  • Data from different channels plotted individually or in 1 single graph

Options & Modules

To check the compatibility of the various options and modules with your designated instrument click here.


Ivium Product Brochure

Ivium Software Brochure